  Power System: Voltage: 85 - 260 Va.c
Frequency: 48 - 62 Hz
Current: > 0.25 A
Fuse: 2 x T250mA 20 x 5 mm
 

 

 
   
  Display: Bargraph Numeric      
100 bar LED 3½ Digit LED    
  Resolution: 2% of range 0.1µm / 1µm      
  Accuracy: ±1Bar or 1% of analogue signal >0.05% reading ±1 count    
  Measurement: Single or Differential    
  Ranges (µm):

±5.0 ±15.0  ±50.0  ±150.0  ±500  ±1500                

   
  Range Selection: 6 way rotary switch on front panel.    
  Zero Adjustment:

Channel: > ±125µm by front panel 10 turn control
Probe: > ±100µm by rear panel multi-turn control

   
  Calibration: Each input can be individually calibrated by recessed control at rear
Range: +/-10%  of nominal.
     
  Connectors: Probe: 2 x 5 pin 240º DIN screw retained.
Data: 9 pin ‘D’ socket
SK1: 15 pin ‘D’ socket
Mains: Switched / Fused IEC power inlet.
 
 
  Probe Inputs: 2 input channels A, B.
Impedance: 10KΩ  +/-2%.
Sensitivity: 50mV / V /mm.
   
  Probe Mix: 3 position slide switch on rear panel or any multiple by resistor via SK1    
  Probe Drive: Voltage: 1.2V rms, nominal
Frequency: 5.0KHz (no load)
Current: 25mA
   
  Tolerance Display: By LED indication: O/S (amber), SIZE (green), U/S (red).
Setting Range: Any point on scale (no crossover)
   
  Outputs: RS232: via 9 way D socket
Probe signals: via SK1
Analogue: via SK1
Tolerance: via SK1 1 TTL standard load
DC power: +/- 5V at 10mA
   
  Environmental: Storage: -10°C to +70°C
Working:  0°C to +40°C
Humidity: 10 - 95% non-condensing
   
  Mechanical: Enclosure: 2mm Mild Steel Painted
Size: 384 h x 130 d x 44 w
Weight: 2.85 Kg
     
  Electromagnetic:
Conformity
EN61326:1998 Class A
EN61010-1:2001